Remove the temperature probe and probe well
Follow these steps to disconnect the probe cable and remove the probe well.
1. On the underside of the probe well, press the spring tab on the SureTemp probe
connector and pull it out of the probe well.
2. Remove the SureTemp probe from the probe well.
3. Grasp the probe well and pull it up to remove it from the monitor.
Connect the NIBP hose
1. Place your thumb and forefinger on the hose connector spring tabs and squeeze
firmly.
2. Align the hose connector with the hose connector port on the bottom of the monitor.
3. Insert the hose connector, pressing firmly until both of the spring tabs click into
place.
Disconnect the NIBP hose
1. Place your thumb and forefinger on the hose connector spring tabs.
Note Always grasp the hose by the connector spring tabs. Do not
pull on the hose itself.
2. Squeeze and pull the spring tabs until the connector releases.
Connect the SpO2 cable
WARNING Patient injury risk. Do not use a damaged sensor or pulse
oximetry cable or a sensor with exposed electrical or optical components.
1. On the bottom of the monitor, align the SpO2 cable connector with the cable
connector port.
2. Insert the cable connector, pressing firmly until the connector is seated.
Disconnect the SpO2 cable
1. Place your thumb and forefinger on the Sp02 cable connector. Do not grasp the
cable.
2. Pull the Sp02 cable connector out of the connector port.
Attach an accessory
CAUTION Accessories attached to this monitor must run on battery
power. Do not use any accessory's external power supply when it is
attached to the monitor.
To attach an accessory to the monitor, follow the Directions for use that
accompanied the accessory.
22 Setup Connex
®