Product Manual
Setup
Supplies and accessories
For a list of all approved supplies and accessories, see
Approved Accessories in the
Appendices.
WARNING Patient injury risk. Clean all accessories, including cables and
tubes, before storing the accessories on the device or cart. This helps
reduce the risk of cross contamination and nosocomial infection. Refer to
'Clean the equipment' in "Maintenance and service" for directions.
Connect the battery
This procedure applies to first-time setup of the monitor. The battery is inserted in the
battery compartment when you receive a new monitor. However, it is not connected.
WARNING Personal injury risk. Improper handling of the battery can lead
to heat generation, smoke, explosion, or fire. Do not short-circuit, crush,
incinerate, or disassemble the battery. Never dispose of batteries in refuse
containers. Always recycle batteries according to national or local
regulations.
WARNING
Use only Welch Allyn approved accessories, and use them according to the
manufacturer’s directions for use. Using unapproved accessories with the
monitor can affect patient and operator safety and can compromise product
performance and accuracy, and void the product warranty.
1. Turn the monitor screen down to access the battery cover.
2. Locate the battery cover, indicated by on the back of the monitor.
3. Using a double-slotted screwdriver, loosen the captive screw at the base of the
battery cover, and then remove the cover.
4. Remove the battery to access the battery connection port on the monitor.
5. Insert the battery connector into the battery connection port on the monitor.
6. Insert the battery into the battery compartment.
7. Replace the battery cover, and then tighten the captive screw at the bottom of the
battery cover.
Note Do not over-tighten the screw.
19