Product Manual
NIBP specifications
Blood pressure determination time
Typical: 15 seconds
Maximum: 150 seconds
Blood pressure accuracy Meets or exceeds ANSI.AAMI SP10:2002 standards for
noninvasive blood pressure accuracy (±5 mmHg mean error,
8 mmHg standard deviation)
Mean Arterial Pressure (MAP) range
The formula used to calculate MAP yields an approximate
value.
Adult: 23 to 230 mmHg (StepBP, SureBP)
Pediatric: 23 to 230 mmHg (StepBP, SureBP)
Neonate: 13 to 110 mmHg (StepBP)
Pulse rate range (using blood pressure determination)
Adult: 30 to 200 bpm (StepBP, SureBP)
Pediatric: 30 to 200 bpm (StepBP, SureBP)
Neonate: 35 to 220 bpm (StepBP)
Pulse rate accuracy (using blood pressure determination) ±5.0% (±3 bpm)
Overpressure cutoff
Adult: 300 mmHg ±15 mmHg
Pediatric: 300 mmHg ±15 mmHg
Neonate: 150 mmHg maximum
SureTemp Plus temperature module specifications
SureTemp Plus temperature module specifications
Temperature range 80°F to 110°F (26.7°C to 43.3°C)
Calibration accuracy ±0.2°F (±0.1°C) (Direct mode)
Braun ThermoScan Pro 6000 specifications
Braun ThermoScan PRO 6000 thermometer specifications (refer to Braun ThermoScan Pro 6000's
directions for use for additional information)
Temperature range 68°F to 108°F (20°C to 42.2°C)
Calibration accuracy
• ±0.4°F (±0.2°C) for temperatures ranging from 95.9°F to 107.6°F (35.5°C to
42°C)
• ±0.5°F (±0.25°C) for temperatures outside of this range
Display resolution 0.1°F or °C
130 Specifications Connex
®