Product Manual
2 Cell battery specifications
1
Hours of use
Acute care continuous 10 minute cycles - 49 patient cycles - BP, temp, SpO2, no radio,
no scanner (Nellcor)
8.22
Acute care continuous 10 minute cycles - 50 patient cycles - BP, temp, SpO2, no radio,
no scanner (Nonin)
8.37
Acute care continuous 10 minute cycles - 49 patient cycles - BP, temp, SpO2, no radio,
no scanner (Masimo)
8.29
Acute care continuous 10 minute cycles - 41 patient cycles - BP, temp, SpO2, radio,
scanner (Nellcor)
6.84
Acute care continuous 10 minute cycles - 41 patient cycles - BP, temp, SpO2, radio,
scanner (Nonin)
6.96
Acute care continuous 10 minute cycles - 41 patient cycles - BP, temp, SpO2, radio,
scanner (Masimo)
6.90
1
Nellcor is the default for these examples.
Nurse Call specifications
Nurse Call connection specifications
Nurse Call 25 V AC or 60 V DC maximum at 1A maximum
NIBP specifications
NIBP specifications
Cuff pressure range Meets or exceeds IEC/ISO 80601-2-30 standards for cuff
pressure range
Systolic range
Adult: 30 to 260 mmHg (StepBP, SureBP)
Pediatric: 30 to 260 mmHg (StepBP, SureBP)
Neonate: 20 to 120 mmHg (StepBP)
Diastolic range
Adult: 20 to 220 mmHg (StepBP, SureBP)
Pediatric: 20 to 220 mmHg (StepBP, SureBP)
Neonate: 10 to 110 mmHg (StepBP)
Cuff Inflation Target
Adult:160 mmHg (StepBP)
Pediatric: 140 mmHg (StepBP)
Neonate: 90 mmHg (StepBP)
Maximum Target Pressure
Adult: 280 mmHg (StepBP, SureBP)
Pediatric: 280 mmHg (StepBP, SureBP)
Neonate: 130 mmHg (StepBP)
Directions for use Specifications 129