Product Manual
Protection classifications, all monitor configurations
Graphical display resolution
Dimensional outline 6.5 in. (W) x 4.1 in. (H) x 0.13 in. (D) (164.9 mm [H] x 103.8
mm [W] x 3.40 mm [D] )
Active area 6.1 in. (W) x 3.4 in. (H) (154.08 mm [W] x 85.92 mm [H])
Resolution 800 x 480 pixels
Pixel arrangement RGB (red, green, blue)
Pixel size 63.2 µm (W) x 179 µm (H)
Luminance 530 cd/m2
Speaker volume
Minimum Output sound pressure 60 dB at 1.0 meter
Alarm and pulse tones
per IEC 60601-1-8
Pulse frequency (f
0
) 150 – 1000 Hz
Number of harmonic components in the range 300 Hz to
4000 Hz
minimum of 4
Effective pulse duration (t
d
) high priority: 75 –200 ms
medium and low priority: 125 – 250 ms
Rise time (t
r
) 10 – 20% of t
d
Fall time
1
(t
f
)
t
f
< t
s
– t
r
Note
The range of the relative sound pressure level of the harmonic components should be between a
minimum of at least 53 dBa and a maximum of at least 80 dBa at the pulse frequency.
1
Prevents overlap of pulses
Battery specifications
2 Cell battery specifications
1
Hours of use
Continuous run time (Nellcor) 5.22
6 patients/hour - 41 patient cycles (Nellcor) 6.83
8 patients/hour - 54 patient cycles (Nellcor) 6.78
8 patients/hour - 55 patient cycles (Nonin) 6.90
128 Specifications Connex
®
Spot Monitor