Product Manual
Message Possible cause Suggested action
Alarm
priority
Clinician ID required to
start intervals.
A Clinician ID is required to start
intervals
Information status message; press OK button
to dismiss.
Information
Patient ID match
required to save data.
A Patient ID match is required to
save data
N/A Information
Patient ID match
required to start
intervals.
A Patient ID match is required to
start intervals
N/A Information
Clinician ID match
required to save data.
A Clinician ID match is required to
save data
N/A Information
Clinician ID match
required to start
intervals.
A Clinician ID match is required to
start intervals
N/A Information
Unable to auto save. The device was unable to auto
save
Information status message; press OK button
to dismiss.
Information
Barcode scan not
accepted.
Barcode scanning is not available
Information status message; press OK button
to dismiss.
Information
Invalid NIBP interval
parameter during
interval capture.
An invalid interval parameter was
detected.
Information status message; press OK button
to dismiss.
Information
Save successful. The auto save is successful in
Office Profile
Information status message; press OK button
to dismiss.
Information
Unsent records: N of M There are unsent records waiting
when the device is powered down
Information status message; press OK button
to dismiss.
Information
Barcode scanning is not
available. Enter patient
information manually.
Barcode scanning is not available.
Enter patient information
manually.
Information status message; press OK button
to dismiss.
Information
Invalid SpO2 interval
parameter during
interval capture.
An invalid interval parameter was
detected.
If SpO2 intervals are enabled and SpO2 clip
was removed, either stop the intervals or
reattach the SpO2 clip. Information status
message; press OK button to dismiss.
Information
Radio messages
Message Possible cause Suggested action
Alarm
priority
Radio not functional.
350001
Deserialization failure. There is a
software communication issue
between the host and the radio
Check for a software update and apply it.
If still present, replace the radio.
Very low
114 Troubleshooting Connex
®
Spot Monitor