Product Manual
Message Possible cause Suggested action
Alarm
priority
Temperature not functional.
3F0824
The battery is over the
maximum voltage
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0833
The sensor not
functional
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0E04
Low battery Recharge the battery. If the problem
persists, check the battery.
Very low
Unable to detect new
temperature. Retry measurement.
No temperature
measurement was
available from the
thermometer at the
time it was docked.
If a measurement should have been
available, retry the measurement. If
problem the persists, replace the module.
Information
Thermometer might be docked
improperly. Check contacts and
connections.
Communication failure
with docked Braun
The thermometer might be docked
improperly. Check contacts and
connections. If problem the persists,
replace the module.
Information
Temperature not functional.
3FFF01
Unrecognized WACP
parameter received
from sensor
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3FFF02
Timeout waiting for
sensor response
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3FFF03
Error deserializing
WACP message
received from sensor
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3FFF04
WACP stack send
message failure
Internal malfunction. If the problem
persists, replace the module.
Very low
Re-dock Braun. 3FFF05 Anti-theft timer expired Re-dock the thermometer after taking a
measurement.
Very low
Patient and clinician data messages
Message Possible cause Suggested action
Alarm
priority
Unable to identify
clinician. No provider
configured at host.
Clinician authentication failure Information status message; press OK button
to dismiss.
Information
Unable to identify
clinician. Security
provider error.
Clinician authentication failure Information status message; press OK button
to dismiss.
Information
Unable to identify
clinician. User not
found.
Clinician authentication failure Information status message; press OK button
to dismiss.
Information
112 Troubleshooting Connex
®