Message Possible cause Suggested action
Alarm
priority
Temperature not functional.
3F0208
The object provided
with the message could
not be deserialized.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0209
The object could not be
serialized.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F020A
The message is
performing a request/
action when the
module state prohibits
the request/action.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F020B
The requested item is
not currently available
due to the module
state.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0503
The factory settings,
and calibration
information is corrupt.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0504
The user settings are
corrupt.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0509
The calibration is not
set.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F050C
The error log is corrupt. Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0516
A hardware
malfunction has been
detected
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0518
The module power rail
is too low.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0519
The module power rail
is too high.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F051A
The reference voltage
circuit was detected to
be under voltage or
unstable.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0821
The ambient
temperature is too high
Verify conditions are less than 104°F or
40°C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
Very low
Temperature not functional.
3F0822
The ambient
temperature is too low
Verify conditions are greater than 50°F or
10°C. If conditions are valid and problem
persists, replace the probe. If the problem
still persists, replace the module.
Very low