Product Manual
Message Possible cause Suggested action Alarm priority
Temperature not functional.
03CE01
Upgrade file directory
could not be accessed
Retry the device update Very low
Direct mode reading timed out Direct mode reading
times out
Direction mode reading times out Information
Tissue contact lost. Lost tissue contact
while attempting to
acquire a temperature
measurement or
acquired measurement
was performed with
limited tissue contact.
Check the tissue contact and try the
measurement again.
Information
Temperature module reset.
03D000
Temperature sensor
reset unexpectedly
None Very low
Braun 6000 messages
Message Possible cause Suggested action
Alarm
priority
Temperature not functional.
3F0105
WACP message CRC
mismatch.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0201
This message is not
implemented by the
module.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0202
This message is not
supported by the
module.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0203
The module has run out
of memory.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0204
No parameter provided
for the specified
message.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0205
The parameter provided
is invalid for the
specified message.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0206
The parameter provided
is outside of the
allowable range for the
specified message.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3F0207
The message requires
an object, but did not
contain one.
Internal malfunction. If the problem
persists, replace the module.
Very low
110 Troubleshooting Connex
®
Spot Monitor