Message Possible cause Suggested action Alarm priority
Temperature not functional.
30825
The temperature
module is under min
battery volts
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30826
The temperature
module battery volts
not set
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30827
The temperature
module predict
algorithm is not set
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30828
The temperature
module ambient temp
is not set
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30829
The temperature
module has a non-
responsive probe. The
thermistor has pulled
away from the tip or
the heater is broken.
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
3082A
The temperature
module is experiencing
bad probe gain
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
3082B
The temperature
module has a bad probe
response value
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
03C800
The temperature
module is not
functional
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
03C900
Unable to deserialize
messages from the
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
03CA00
Unsupported message
received from the
Temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
03CB00
Unable to send
message to the
Temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
03CC00
Temperature module
communication times
out
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
03CD00
Failed to upgrade the
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
03CE00
Unable to read PIM file Retry the device update. Very low