Product Manual

Message Possible cause Suggested action Alarm priority
indicates on when
turned off.
Unable to detect new
temperature. Retry measurement.
30819
The temperature
module heater
indicates off when
turned on.
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
3081A
The temperature
module HTR_Q is on
and HTRC is off but still
has voltage.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3081B
The temperature
module HTR_Q is tri-
stated with HTRC
enabled and has heater
power.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3081C
The temperature
module turned Q&C on
and the heater voltage
is not high enough.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3081D
The temperature
module heater
hardware failsafe
should have turned off
but did not.
Internal malfunction. If the problem
persists, replace the module.
Very low
Replace temperature probe.
3081E
The temperature
module probe is above
112°F or 43.3°C.
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Replace temperature probe.
3081F
The temperature
module has excessive
heater energy
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
30820
The temperature
module host interface
error
Internal malfunction. If the problem
persists, replace the module.
Very low
Ambient temperature out of
range. Clear to retry. 30821
The temperature
module over ambient
45 °C temp
Verify conditions are less than 104 °F or
40 °C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
Very low
Ambient temperature out of
range. Clear to retry. 30822
The temperature
module under ambient
temp
Verify conditions are greater than 50 F or
10°C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
Very low
Temperature not functional.
30823
The temperature
module has an invalid
SureTemp algorithm
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30824
The temperature
module is over max
battery volts
Internal malfunction. If the problem
persists, replace the module.
Very low
108 Troubleshooting Connex
®
Spot Monitor