Product Manual

Message Possible cause Suggested action Alarm priority
detection mechanism
detected an error
Temperature not functional.
3080D
The temperature
module log error
detection mechanism
detected an error
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
3080E
The temperature
module calibration error
detection mechanism
detected an error
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Connect temperature probe.
3080F
The temperature
module detected no
probe connected
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Replace temperature probe.
30810
The temperature
module cannot read the
probe EEPROM
correctly or the probe
left the factory without
being tested.
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
30811
The temperature
module has an invalid
events index
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30812
There is a problem
reading the
temperature module
EEPROM or saving to
the monitor EEPROM in
biotech mode.
Internal malfunction. If the problem
persists, replace the module.
Very low
Replace temperature probe. Code
30813
The temperature
module has a problem
reading the probe
EEPROM.
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Temperature not functional.
30814
The temperature
module TEMP CONFIG
ACQUIRE FAILURE
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30815
The temperature
module TEMP CONFIG
RELEASE FAILURE
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30816
The temperature
module TEMP CONFIG
INVALID PTR FAILURE
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30817
The temperature
module internal error.
EEPROM not initialized
Internal malfunction. If the problem
persists, replace the module.
Very low
Unable to detect new
temperature. Retry measurement.
30818
The temperature
module heater
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
Directions for use Troubleshooting 107