Message Possible cause Suggested action Alarm priority
ambient or patient
limits.
Temperature not functional.
30803
The temperature
module internal
calibration resistor
(RCAL) on the board is
damaged or
contaminated (pulse
too long).
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30804
The temperature
module internal
calibration resistor
(RCAL) on the board is
damaged or
contaminated (pulse
too short).
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30805
The temperature
module internal circuit
validation resistor (PTB)
on the board is
damaged (value is
over).
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30806
The temperature
module internal circuit
validation resistor (PTB)
on the board is
damaged (value is
under).
Internal malfunction. If the problem
persists, replace the module.
Very low
Ambient temperature out of
range. Clear to retry. 30807
The temperature
module A/D
measurement timed out
Verify conditions are greater than 50 °F or
10 °C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
Very low
Replace temperature probe.
30808
The temperature
module probe was not
characterized/
calibrated
Probe malfunction. Replace the probe. If
the problem persists, replace the module.
Very low
Insert correct color-coded probe
well. 30809
The temperature
module is missing the
probe well
Insert the probe well Very low
Temperature not functional.
3080A
The temperature
module has a problem
saving to the monitor
EEPROM in biotech
mode
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3080B
The temperature
module error detection
mechanism detected an
error
Internal malfunction. If the problem
persists, replace the module.
Very low
Replace temperature probe.
3080C
The temperature
module probe error
Probe malfunction. Replace probe. If the
problem persists, replace the module.
Very low
106 Troubleshooting Connex
®