Message Possible cause Suggested action Alarm priority
module state prohibits
the request/action.
Temperature not functional.
3020B
The temperature
module requested item
is not currently
available due to the
module state.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30503
The temperature
module factory
settings, and
calibration information
is corrupt.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30504
The temperature
module user settings
are corrupt.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30509
The temperature
module calibration is
not set.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3050C
The temperature
module error log is
corrupt.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30516
A hardware
malfunction on the
temperature module
has been detected.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30518
The temperature
module power rail is
too low.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30519
The temperature
module power rail is
too high.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3051A
The temperature
module reference
voltage circuit was
detected to be under
voltage or unstable.
Internal malfunction. If the problem
persists, replace the module.
Very low
Ambient temperature out of
range. Clear to retry. 30801
The temperature
module measurement is
below the allowable
temperature values and
beyond the low
ambient or patient
limits.
Verify conditions are greater than 50 °F or
10 °C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
Very low
Ambient temperature out of
range. Clear to retry. 30802
The temperature
module measurement is
above the allowable
temperature values and
beyond the high
Verify conditions are less than 104 °F or
40 °C. If conditions are valid and the
problem persists, replace the probe. If the
problem still persists, replace the module.
Very low