Product Manual

Message Possible cause Suggested action Alarm priority
tester. If the message persists,
replace the module.
1
Pulse search is a normal part of operation and does not have an associated corrective action.
Temperature messages
SureTemp messages
Message Possible cause Suggested action Alarm priority
Temperature not functional.
30105
WACP message CRC
mismatch on
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30201
This message is not
implemented by the
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30202
This message is not
supported by the
temperature module
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30203
The temperature
module has run out of
memory.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30205
The temperature
module received a
invalid parameter
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30206
The parameter provided
by the temperature
module is outside of
the allowable range for
the specified message.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30207
The temperature
module message
requires an object, but
did not contain one.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30208
The temperature
module object provided
with the message could
not be deserialized.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
30209
The temperature
module object could not
be serialized.
Internal malfunction. If the problem
persists, replace the module.
Very low
Temperature not functional.
3020A
The temperature
module message is
performing a request/
action when the
Internal malfunction. If the problem
persists, replace the module.
Very low
104 Troubleshooting Connex
®
Spot Monitor