Product Manual
Message Possible cause Suggested action Alarm priority
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
SpO2 rebooting. 042f00 There is a problem with the SpO2
positive detector voltage
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 043000 There is a problem with the SpO2
negative detector voltage
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 043100 There is a problem with the SpO2
positive LED Voltage
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 043200 There is a problem with the SpO2
LED drive voltage
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 043300 There is a problem with the SpO2
positive preamp voltage
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
Very low
100 Troubleshooting Connex
®
Spot Monitor