Message Possible cause Suggested action Alarm priority
Replace the SpO2
sensor. 040f00
The SpO2 sensor is defective Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Replace the SpO2
sensor. Replace the
SpO2 cable. 041000
There was a fault with the SpO2
sensor and cable.
Check the sensor and cable
connection. If the problem is still
present, replace the SpO2 sensor. If
the problem is still present, replace
the cable. If the problem is still
present, verify the module
functionality by replacing the sensor
with an applicable SpO2 tester. If the
message persists replace the
module.
Very low
Sensor not connected.
Clear to retry. 041100
An adhesive SpO2 sensor is not
connected
Connect SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
The sensor has expired.
Replace the SpO2
sensor. 041200
The adhesive SpO2 sensor is
expired
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Incompatible sensor.
Replace the SpO2
sensor. 041300
The adhesive SpO2 sensor is
incompatible
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Incompatible sensor.
Replace the SpO2
sensor. 041400
The adhesive SpO2 sensor is not
recognized
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Replace the SpO2
sensor. 041500
The adhesive SpO2 sensor is
defective
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
Very low