Product Manual
Message Possible cause Suggested action Alarm priority
tester. If the message persists,
replace the module.
Replace the SpO2 cable.
040800
The SpO2 cable is incompatible
with the monitor
Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Very low
Replace the SpO2 cable.
040900
The SpO2 cable is not recognized
by the monitor
Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Very low
Replace the SpO2 cable.
040a00
The SpO2 cable is defective Replace the SpO2 cable. If the
problem is still present, verify
module functionality by replacing the
sensor with an applicable SpO2
tester. If the message persists,
replace the module.
Very low
Sensor not connected.
Clear to retry. 040b00
The SpO2 sensor is not connected
to the monitor
Connect SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
The sensor has expired.
Replace the SpO2
sensor. 040c00
The SpO2 sensor has expired Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Incompatible sensor.
Replace the SpO2
sensor. 040d00
The SpO2 sensor is not recognized
by the monitor
Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
Incompatible sensor.
Replace the SpO2
sensor. 040e00
The SpO2 sensor is not recognized Replace SpO2 sensor. If the problem
is still present, replace the SpO2
cable. If the problem is still present,
verify the module functionality by
replacing the sensor with an
applicable SpO2 tester. If the
message persists, replace the
module.
Very low
94 Troubleshooting Connex
®
Spot Monitor