Product Manual
Table Of Contents

6 WELCH ALLYN UNIVERSAL DESK CHARGER SERVICE MANUAL 711380 REV. D
3.1.1 Green LED not on.
Use a known good power cord.
Replace the power cord if this fixes
 the problem.
Check Fuses
If fuses are OK, then replace printed
circuit board.
3.1.2 Amber charge LED not on.
Insert a Welch Allyn rechargeable
handle in charging well or T17814
repair test handle. If the amber
charge light still does not come on,
then replace printed circuit board.
3.1 Trouble Shooting