Product Manual

Directions for Use Fiber Optic Laryngoscope Blades and Handles 3
Fiber Optic Laryngoscope Handle
Battery handles consist of two sections: the main handle and the lamp holder cartridge
assembly (see Figure). Prior to cleaning and disinfection, disassemble the handle
following the procedure outlined in the Repair/Maintenance section. The main handle will
withstand the same cold soak solutions and autoclave ranges outlined in the
Laryngoscope Blades Section. However, the lamp holder cartridge assembly and
batteries must be removed prior to disinfection/sterilization. See figure 3.
Cleaning Procedure
Main Handle Section
Remove the batteries and lamp cartridge as described in the Repair/Maintenance section.
The main handle section may then be cleaned with a mild detergent and warm water
solution. The main handle section may also be soaked in an enzymatic detergent
following the manufacturers instructions.
Lamp Cartridge
The lamp cartridge may be wiped clean with a mild detergent and water solution. DO
NOT ALLOW SOLUTION TO ENTER INTO CARTRIDGE.
Use of instrument in presence of intense magnetic fields should present no problems.
Disinfection/Sterilization
Main Handle Section
Remove batteries and lamp cartridge before subjecting handle to any of the following
processes.
After removing the batteries and lamp cartridge, the main handle can be soaked in a 14
day (2.4 - 2.6%) glutaraldehyde solution following manufacturer’s instructions for
exposure time and temperature.
Steam autoclave is appropriate. Do not exceed temperature of 280°F (138°C) and
pressure of 28 psi. DO NOT FLASH AUTOCLAVE.
Ethylene Oxide is appropriate with gas concentrations of 10% - 100%. Do not exceed
131°F (55°C). Maximum 4 hours exposure time. Aerate 12-16 hours at 120° F (49°C)
following process.
WARNING Only trained personnel shall use a laryngoscope for intubation.