Note As shown here, an information message prompts you to
select a pediatric age range when you monitor IPI on
pediatric patients.
8. Touch OK.
The Home tab appears.
Note Pediatric patient subtypes appear on the Patient Summary
tab and on the Home tab as a label in the IPI frame, but they
do not appear in the Patient frame.
Configure IPI alarms
Follow these steps to set alarm limits for IPI.
1.
Verify that you are using the Continuous Monitoring profile.
2. Touch the Alarms tab.
3. Touch the IPI vertical tab.
4. Ensure that the Limits control is set to ON.
Note If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Adjust the IPI limits setting as desired. Enter the desired lower alarm limit for IPI
using the up/down arrow keys or the keypad.
6. Touch the Home tab.
The new alarm settings take effect immediately.
Acoustic respiration rate (RRa)
Monitors configured with Masimo rainbow Acoustic Monitoring technology continuously
and noninvasively measure respiration rate based on respiratory sounds produced in the
upper airway. Respiratory sounds are detected through a sensor applied to the patient's
neck.
Directions for use
Patient monitoring 89