Product Manual
4. Select the appropriate sampling line based on the following considerations:
• Whether the patient is intubated or non-intubated
• Whether the patient is on mechanical ventilation
• Duration of use
• Patient's size and weight
5. Connect the sampling line to the monitor.
a. Slide open the door that protects the sampling line port on the monitor.
b. Connect the sampling line.
Note The CO2 pump starts as soon as the pump sensor
recognizes the sampling line. This can occur before you
completely connect the sampling line.
c. Turn the connector completely to the right until tight.
An etCO2 sampling line dialog appears on the screen to reinforce this instruction
(unless the dialog is disabled in Advanced settings).
WARNING You must connect the sampling line correctly.
Ensure that the connector is tight to accurately monitor CO2.
Note This secure connection prevents gases from leaking at the
connection point during measurement and is essential to
ensure accurate measurements.
6. If the etCO2 sampling line dialog appears on your screen, watch the animation and
confirm that you have connected the sampling line correctly, then touch OK to
dismiss the dialog.
7. Connect the sampling line to the patient as described in the directions for use
supplied with the sampling line.
8. Perform the following checks:
a. Check that CO2 values appear on the monitor.
b. Check the patient type. Make sure it is correct to ensure accurate sampling.
c. Touch the alarm control to ensure the alarms are set appropriately.
84 Patient monitoring Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™