The alarm clears.
You reset (pause or turn off) the alarm tone.
For paused alarms, Nurse Call notification returns after the pause interval if the condition
that caused the alarm persists.
Nurse Call thresholds are set in the Advanced Settings.
To connect the monitor to a Nurse Call system, you must have a cable that has been
adapted to your Nurse Call system (REF 6000-NC), rated 25V AC or 60V DC maximum at
1A maximum. For ordering information, see Approved Accessories on the CD.
WARNING The monitor is the primary alarming source for the patient, and
any external system (such as a Nurse Call or software system) is a backup
alarm source. The external system is only as reliable as its components
(e.g., wiring, hardware, light bulbs, connections) and can be relied on only
as a backup alarming device.
80 Alarms Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™