Product Manual
Icons in parameter frames
Icon Name and status
Alarm audio paused.
The audio tone is paused. The icon remains until the paused time counts down to 0.
Icons in the Device Status area
The icons in the Device Status area are black and white, but the background area
changes colors to indicate the alarm priority. Messages accompany these icons. These
icons can be controls or status indicators.
Icons in the Device Status area
Icon Name and status
Alarm active.
One or more alarms are active. Touch this icon to reset (pause or turn off) the alarm tone.
WARNING Do not pause or turn off an audible alarm if patient safety might be
compromised.
Alarm audio off.
Audio signals are disabled, but alarm limits and visual alarm signals remain active.
If the patient is being monitored continuously, maintain a clear line of sight with the monitor while
audio alarms are disabled.
Multiple alarms toggle.
Touch this icon to cycle through the messages for each active alarm.
Alarm audio paused.
The audio tone is paused. The icon remains until the paused time counts down to 0. Touch this
icon to start the user-configured pause interval (set in the Advanced tab).
Reset (pause or turn off) audio alarms
WARNING Do not pause or turn off an audible alarm if patient safety
might be compromised.
Audio alarm characteristics
•
After you reset an audio alarm, some tones do not return, but others return after a
pause interval if the condition that caused the alarm persists.
• If a new alarm condition occurs during a pause interval, a new audio tone occurs.
• If an audio alarm is not paused or turned off after a period of time, a buzzer
accompanies the tone.
Pause or turn off an audio alarm
72
Alarms Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™