Product Manual
Type Priority Color Alarm audio tone
• Technical alarms that don't impact patient
safety
• Some calibration information messages
Very low Cyan 2-pulse tone or 1-pulse tone at
5-minute intervals
Alarm notification locations on the monitor
WARNING Whenever possible, do not rely on visual alarm notifications
alone while monitoring patients. If you must rely on visual alarm
notifications, maintain a clear line of sight with the monitor. For audio alarm
notifications, set the volume as needed considering the environment and
ambient noise levels. Verify that the alarm is audible to a clinician working
at the maximum distance from the monitor.
LED light bar
The light bar on the handle of the monitor illuminates as follows:
• Flashing red for high-priority alarms
• Flashing amber for medium-priority alarms
• Constant amber for low-priority and very low-priority alarms
The light bar darkens when the alarm tone is reset.
Home tab
Home tab notifications
Notification Description
Device Status area The area changes color and displays a message with an accompanying status icon or
button. If the alarm tone is in a pause interval, a timer countdown appears.
If multiple alarms are active, the device provides options to manually toggle or autoscroll
through the alarms. You can cycle through each alarm message in sequence of priority by
touching the multiple alarm toggle. Alternatively, the autoscroll option displays each
alarm message in sequence of priority for approximately 4 seconds, and then loops back
70 Alarms Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™