Manage patient records (Continuous Monitoring profile)
In the Continuous Monitoring profile, the Review tab enables access to a trend table of
all readings for the currently selected patient. You can configure the time intervals
displayed, and you can scroll through the last 24 hours of patient measurements. Trend
data can be printed from this tab.
1. Touch the Review tab.
Note Measurements that triggered a physiological alarm are
highlighted on this tab.
2. Modify the trend interval (the progression of time displayed horizontially in the table)
by using the View drop-down menu. Trend interval options are 1, 5, 15, 30, and 60
minutes; 2, 4, or 8 hours; and alarms only. The default trend interval is 1 minute.
Note By modifying the trend interval, you can view additional
"events" (e.g., physiological alarms, manual NIBP or
temperature measurements, Manual parameter entries)
between the currently configured time intervals.
3. Use the scroll bar on the right to view patient measurements for any parameters that
do not fit in the display area.
4. Touch the page forward and page backward controls in the table header
to view additional measurements for this patient. The most recent measurements
appear on the right of the table, with older measurements to the left.
5. Touch Print to print the patient's record.
A Print options dialog appears.
6. Select the desired Timespan and touch Print.
Note If no patient is selected and continuous monitoring is not active, all table
cells in the Review tab remain blank.
Note In the Continuous Monitoring profile, you cannot delete saved patient
measurements. Patient measurements older than 24 hours are
automatically deleted from the Review tab.