Product Manual
2. Touch Standby.
A "Standby" dialog appears, indicating that you have paused continuous monitoring.
Controls on this screen provide options to resume or end monitoring. A countdown
timer displays the time remaining before continuous monitoring resumes.
Note You can configure the Standby Mode timeout in Advanced
settings.
Resume continuous monitoring
You can resume continuous monitoring after a standby when patient sensors are
connected to a patient. The status of the standby timer determines your next step.
Standby time remaining
If the standby time limit has not expired (time remains on the countdown timer), resume
monitoring as follows:
1. Reconnect patient sensors to the patient as needed.
2. Touch Resume monitoring.
The Home tab appears and continuous monitoring resumes.
Standby time limit expired (patient sensors connected)
If the standby time limit has expired and you have already reconnected patient sensors
to the patient and/or the device, the Home tab appears and continuous monitoring
automatically resumes.
Standby time limit expired (patient sensors disconnected)
If the standby time limit runs out before you have reconnected patient sensors to the
patient, the monitor returns to the Home tab, and the associated alarms occur.
Reconnect patient sensors to the patient and/or device and acknowledge alarm and
information messages as needed.
Continuous monitoring resumes.
End continuous monitoring
When you no longer need to monitor a patient continuously, follow these steps to end
monitoring:
1.
Touch the Settings tab.
The Monitor tab appears.
2. Touch End monitoring.
An "End monitoring" dialog appears with the following options:
• New patient – Clears patient and location context and review data after you
remove patient sensors