Product Manual

Continuous patient monitoring
Your device must be configured with the Continuous Monitoring profile to perform
continuous patient monitoring.
Key terms
Patient context Condition in which a patient ID and patient type have been selected on a device.
Location context Condition in which a room or a room and bed have been selected on a device.
You have the option to continuously monitor patients on the device alone (primary
monitoring) or to stream continuous patient data to a central station as well (secondary
monitoring).
If Continuous Monitoring has been configured as the default profile, then the device
starts up in this profile. If the Continuous Monitoring profile is active, you can proceed
with an existing patient and/or location context, or you can assign a patient and location.
Activate the Continuous Monitoring profile
If the Continuous Monitoring profile is not active, follow these steps:
1.
Touch the Settings tab.
2. Touch the Device tab.
The Profile vertical tab appears.
3. Select Continuous Monitoring.
If you have stored patient context or patient measurement data, a confirmation
dialog appears with the following message: "Changing to the Continuous Monitoring
profile deletes all measurement data from the device. Continue anyway?"
4. Touch Yes to confirm or No to cancel the profile change.
Note Selecting No enables you to send captured vitals data on the
device before changing profiles. After you complete that
task, if applicable, repeat steps 1–4.
5. After confirming the profile change, touch the Home tab.
You can now start monitoring a new patient in the Continuous Monitoring profile.
Pause continuous monitoring (Standby mode)
You can temporarily pause continuous monitoring and retain patient data on the device
when your patient needs to ambulate, use the restroom, or go off the unit for a test.
Note
If a low-battery alarm occurs on the device, Standby mode is not available.
Note While in standby mode, the device retains previously stored patient data,
but it does not display this data or store additional patient measurements
until you exit standby mode.
1. Touch the Settings tab.
The Monitor tab appears.
52 Connex CS Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™