Product Manual
mode from the active state. Settings for this mode can be changed in the Advanced
Settings Display tab.
Battery-powered accessories connected to the monitor continue to charge while the
monitor is in this mode and connected to AC power.
Note The monitor will not enter the Display power-saving mode while an alarm
condition is active or when NIBP intervals are in progress. In addition, the
monitor will exit this mode if an alarm occurs.
The following actions will return the monitor display to the active state:
• Touch the screen
• Remove the temperature probe from the probe well
• Attach the SpO2 sensor to a patient
• Attach a sampling line to a patient
•
Press
Power up the monitor
The monitor runs a brief diagnostic self-test each time it powers up.
WARNING Equipment failure risk. The monitor includes a fan that
circulates air through the device. If the fan does not run when you power
up the device, remove it from use and inform qualified service personnel
immediately. Do not use the monitor until the problem is corrected.
WARNING To ensure patient safety, listen for two audible indicators (a
piezo beeper and a speaker tone) and watch for visual alerts at power-up at
least once daily. Correct any system errors before using the monitor. In
addition to the audible indicators, the monitor LED light bar illuminates to
alert you of alarms. Amber indicates a low-level alarm. Flashing amber
indicates a medium-level alarm. Flashing red indicates a high-level alarm.
WARNING Always observe the monitor during power-up. If any display
fails to illuminate properly, or if a system fault code or message displays,
inform qualified service personnel immediately, or call your nearest Welch
Allyn Customer Service or Technical Support facility. Do not use the
monitor until the problem is corrected.
CAUTION Always use the monitor with an adequately charged and
properly functioning battery.
CAUTION During continuous monitoring, keep the monitor connected to
AC power at all times.
CAUTION Use only a Class I (grounded) AC power cord for powering this
monitor.
1. Press to power up the monitor.
Following a successful self-test, the monitor displays the Welch Allyn logo, the LED
light bar (located on the handle) flashes, and a power-up tone sounds. The startup
screen then appears with the following banner across the bottom.
32 Startup Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™