Note Labels appear on both the top and bottom of the dual connector as well as
on the patient cables to help ensure proper connection of the patient
cables to the dual cable.
Note Typically, a clinician will connect the single-use RRa sensor to the RRa
patient cable at the start of acoustic respiration monitoring. See the sensor
manufacturer’s directions for use for more information. Also see the
Acoustic Respiration rate (RRa) section of this directions for use.
Disconnect the SpO2 cable or the SpO2/RRa dual cable
1. Place your thumb and forefinger on the cable connector.
Note Always grasp the cable by the connector. Do not pull on the
cable itself.
2. Squeeze the side tabs until the connector releases.
3. Pull the connector away from the connector port.
Attach a USB accessory
CAUTION Accessories attached to this monitor must run on battery
power. Do not use any accessory's external power supply when it is
attached to the monitor.
1. On the rear of the monitor, loosen the screw on the USB door and open it.
Note On some monitor stands, the monitor mounting bracket
partially overlaps the USB door. If you observe this overlap,
loosen the screw on the monitor mounting bracket and shift
the monitor forward just enough on the bracket to open the
USB door, then slide the monitor back on the mounting
bracket.
2. Attach each accessory's USB cable into an unused USB port on the monitor. See the
accessory's directions for use for any special instructions.
CAUTION Connect cables in a manner that minimizes
entangling.
Directions for use Setup 27