Product Manual
Alarm limits Upper limit range of entry Lower limit range of entry
IPI
None Adult, Pediatric: 1 to 10. Factory default: 4.
RR
Adult: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Pediatric: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Neonate: 5 to 150 breaths per minute. Factory
default: 80 breaths per minute.
Adult: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Pediatric: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Neonate: 0 to 145 breaths per minute. Factory
default: 12 breaths per minute.
RRa
Adult: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Adult: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Environmental specifications
Operating temperature 50°F to 104°F (10°C to 40°C)
Transport/storage temperature -4°F to 122°F (-20°C to 50°C)
Operating altitude -557 to 10,000 ft. (-170 m to 3,048 m)
Operating humidity 15 to 95% noncondensing
Transport/storage humidity 15% to 95% noncondensing
Monitor radio
The monitor's radio operates on Welch Allyn FlexNet™ or other 802.11 networks.
Wireless network interface
IEEE 802.11 b/g, 802.11a
Frequency
802.11 b/g: 2.402 GHz to 2.4835 GHz
802.11a: 5.125 GHz to 5.875 GHz
Channels
Up to 14 in 802.11b/g, up to 24 in 802.11a; country-dependent
Security/encryption/authentication
WPA2/AES (either EAP or PSK authentication)
Antenna
Internal multiband PIFA
Wireless data rates
802.11b: 1Mbps or higher during vitals transmission only
802.11a/g: 6Mbps or higher during vitals transmission only
(approximately 2 seconds per reading)
Agency approvals
US: FCC Part 15, Class B; C/UL; CE; 47 CFR Part 2.1093, 15.207, 15.209, 15.247,
15.407; FCC OET Bulletin 65C; FIPS 140-2 Level 1 (Connex VSM 6500 and 6800
series, and Connex IWS 8500 series)
Europe: CE; EN 50371; EN/ETSI 300 328 V1.7.1, 301 489-1 V1.6.1, 301 489-17
V1.2.1, 301 893 V1.4.1
Directions for use Specifications 173