Product Manual

Ethernet messages
Message Possible cause Suggested action
Network not found; check network
cable connection.
A network cable is unplugged Check the network cable
connection. If problem persists,
call for service.
A network connection is broken elsewhere
USB messages
Message Possible cause Suggested action
USB Communication failure. Call
for service.
An internal or external device is connected but
failed enumeration
Power down and restart.
Check USB connections.
If problem persists, call for
service.
External device not licensed for
use.
A license for an external device (e.g., barcode
scanner) has not been activated
Disconnect the unlicensed
device.
Call for service to obtain the
authorization code from Welch
Allyn to activate the license.
External device not recognized. An unrecognized external device is connected Disconnect the unrecognized
device.
Incompatible Welch Allyn device. A communication protocol failure has occurred Call for service.
USB accessory disconnected. The USB cable between an external device and
the monitor is disconnected
Confirm that the USB cable is
connected to the device and the
monitor.
System messages
Message Possible cause Suggested action
Set date and time. The date or time is not set Set the date and time.
The date or time is not set properly Reset the date or time.
Device shutdown is not available
at this time.
Device cannot perform an immediate shutdown
Touch OK, wait, and retry.
Advanced settings unavailable. Sensors are taking measurements Stop continuous measurements.
A physiological alarm condition is active Respond to or reset the alarm.
Spot Check measurements have not been saved Save the measurements.
Unexpected restart occurred. Call
for service.
A system error caused the monitor to restart. Call for service.
158 Troubleshooting Welch Allyn Connex
®