Product Manual

Message Possible cause Suggested action
Reattach the SpO2 sensor to the
patient's finger.
Attach SpO2 sensor to monitor. The sensor was not detected Check the sensor connection.
Replace the SpO2 sensor.
Replace the SpO2 sensor. The SpO2 sensor is faulty or expired Replace the SpO2 sensor.
No SpO2 sensor is connected Connect an SpO2 sensor.
The cable is faulty or expired Replace the cable.
Replace the SpO2 cable. The cable is faulty or expired Replace the cable.
Low SpO2 signal quality. Check
sensor.
Low SpHb signal quality. Check
sensor.
Low perfusion. Check sensor.
Poor sensor placement on the patient Remove the sensor from the
patient and reapply.
The patient cable or sensor is defective Call service to test or replace the
module.
The SpO2 module is faulty Remove the sensor from the
patient and reapply.
SpO2 mode only. Check sensor or
cable.
The sensor is operating as an SpO2-only sensor
because it failed to calibrate properly
Reattach the cable to the
monitor.
Remove the sensor from the
patient and reapply.
SpO2 sensor expires in….
Note This message only appears
on devices configured with SpHb.
The SpO2 sensor will expire soon Replace the SpO2 sensor.
Temperature messages
Message Possible cause Suggested action
Connect temperature probe. No probe is connected Connect a temperature probe and
retry.
The probe is faulty Replace the temperature probe.
The temperature module returned a connect
probe message
Connect a temperature probe and
retry. If a probe is already
connected, replace the probe.
Insert correct color-coded probe
well.
The probe well is missing Insert a temperature probe well.
Replace temperature probe. The probe is faulty Replace the temperature probe.
Temperature not functional. Call
for service.
A module error occurred Call for service.
Directions for use Troubleshooting 155