Product Manual
RRa messages
Message Possible cause Suggested action
Respiratory freshness timeout
expired.
Excessive ambient or environmental noise Reduce the level of noise in the
room.
Poor sensor placement on the patient Remove the sensor from the
patient and reapply.
The patient cable or sensor is defective
Replace the patient cable or
sensor.
Replace RRa probe. The sensor is faulty Replace the sensor.
No sensor is connected Connect the sensor.
The cable is faulty Replace the cable.
RRa patient interference detected. Poor sensor placement on the patient Remove the sensor from the
patient and reapply.
The patient is talking
Ask the patient to limit talking.
The patient is snoring Gently adjust the patient's
position to reduce snoring.
Awaken and ask the patient to
adjust their position to reduce
snoring.
RRa background interference
detected.
Excessive ambient or environmental noise Reduce the level of noise in the
room.
NIBP messages
Message Possible cause Suggested action
NIBP air leak; check cuff and tubing
connections.
The NIBP module has an air leak
Check the cuff, tubing, and
connections for leaks. If no leaks
are found, clear the alarm and
retry NIBP.
If the message reappears, call for
service to replace the NIBP
module.
NIBP not functional. Call for
service.
A module error occurred Call for service.
The ambient temperature is out of range Use the monitor in the specified
temperature range.
Unable to determine NIBP; check
connections; limit patient
movement.
The NIBP module detected a motion artifact Check connections; limit patient
movement.
Clear the alarm and retry NIBP.
Directions for use Troubleshooting 153