Product Manual
• To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify server settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Network tab.
3. Touch the Server tab.
4. Specify settings.
Setting Action/Description
Obtain server IP information automatically Enable the monitor to automatically obtain the server IP
information via the network.
IP address Touch and enter the IP address of the server that is
used for patient data communication. The range of entry
for each field is 0 to 255.
Port Touch and enter the port number associated with
the server IP address. The range of entry is 0 to 65535.
UDP broadcast port Touch and enter the port number that is used to
automatically obtain server IP information. The range of
entry is 0 to 65535.
Test Touch Test to test the connection to the configured
server.
5. Do one of the following:
• To continue in the Advanced Settings, touch another tab.
• To exit the Advanced Settings and return to the Home tab, touch Exit.
Service
The Service tab presents numerous settings and controls typically accessed by
authorized service or biomedical engineering personnel to configure, maintain, test, and
update the device. For example, the Service tab enables authorized users to save device
configurations to a USB flash drive and then load saved configurations to other devices.
Systems and devices configured with the PartnerConnect™ service feature also have
access to remote diagnostics, troubleshooting, and software upgrades.
For a description of service-related advanced settings, see the service manual for this
product.
Directions for use
Advanced settings 149