4. Specify settings.
Setting Action/Description
Default view Select primary and secondary views.
Select Display MAP to display mean arterial pressure
(MAP) in the NIBP frame on the Home tab.
If Display MAP is selected, specify which numerics are
primary in the NIBP frame. On the Home tab, clinicians
can touch the NIBP frame to toggle between views.
Unit of measure Select the NIBP unit of measure for display.
Tube type Select the number of tubes that are connected to the
NIBP cuff that is used with this monitor. If you select 1
tube, the only algorithm available for selection is Step.
Algorithm default Select the default algorithm used to determine NIBP
measurements.
Cuff inflation target If you select the Step algorithm, touch and enter a
default cuff inflation target for each type of patient. In
the Patients tab on the Summary tab, clinicians can
change the CITs from the default CITs that you set here.
Allow interval program changes Select to enable user to modify interval program
selections in the Settings > Setup > Interval tab.
5. Do one of the following:
• To continue in the Advanced Settings, touch another tab.
• To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced temperature settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab will appear.
2. Touch the Parameters tab.
3. Touch the Temperature tab.
4. Specify settings.
Setting Action/Description
Unit of measure Select primary units of measure for the temperature
display on the Home tab
Display temperature conversion Select this to display primary units of measure and
secondary units of measure for the temperature display
on the Home tab.
Default SureTemp Plus site Select the default site for SureTemp measurements. The
default site applies when clinicians power up the
monitor and each time clinicians remove the temperature
probe from the well.
Select Last Site to set the default to the site selected
for the last measurement.
Directions for use Advanced settings 143