Product Manual

Last calibration Displays date (XX/XX/XXXX) and time (00:00:00) of last
calibration.
Annual calibration Displays anniversary date (XX/XX) of annual calibration.
For more calibration information, see the Service manual.
5. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced tabs and return to the Home tab, touch Exit.
Specify advanced SpHb settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the SpHb tab.
4. Specify settings.
Setting Action/Description
Default view Select a numeric view or a waveform view as the
primary SpHb display on the Home tab.
Unit of measure Select the primary unit of measure for the SpHb display
on the Home tab.
Default averaging Select the default moving window of time used by the
parameter to calculate the SpHb value and update the
display: short (approximately 1 minute), medium
(approximately 3 minutes), or long (approximately 6
minutes).
Reference Select arterial or venous as the calibrated reference
source.
5. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced tabs and return to the Home tab, touch Exit.
Specify advanced NIBP settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the NIBP tab.
142 Advanced settings Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™