Product Manual
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the Pulse rate tab.
4. Specify settings.
Setting Action/Description
Display source Select this to show the source of pulse rate
measurements (NIBP or SpO2) on the Home tab.
5. Do one of the following:
• To continue in the Advanced Settings, touch another tab.
• To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced etCO2 settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Parameters tab.
3. Touch the etCO2 tab.
4. Specify settings.
Setting Action/Description
Default view Select a numeric view or a waveform view as the
primary etCO2 display on the Home tab.
Display FiCO2 Select to display FiCO2 on the Home tab.
Waveform scale default Select the default waveform scale in etCO2
measurements.
Unit of measure Select primary units of measure for the etCO2 display on
the Home tab.
Check calibration Select to start CO2 calibration check.
Calibrate Select to start CO2 calibration.
Sweep speed default Select the default waveform sweep speed for the CO2
waveform view on the Home tab.
Enable sampling line dialog Select to enable the sampling line dialog to be displayed
when connecting a sampling line to the device.
BTPS compensation Select to enable automatic adjustments for BTPS (body
temperature pressure saturated) to improve the accuracy
of CO2 measurements.
Calibration due hours Displays time remaining until CO2 sensor calibration is
required.
Maintenance due hours Displays time remaining until CO2 periodic sensor
maintenance is required.
Directions for use Advanced settings 141