Product Manual

Display power saver Specify the required period of monitor inactivity before
the display turns off.
Clinician interactions, new vital sign measurements, or
alarm conditions automatically turn on the display.
Device power down Specify the required period of monitor inactivity before
the monitor turns off.
4. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced device settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Device tab.
3. Specify settings.
Option Description
Location ID Touch and enter up to 20 alphanumeric
characters.
Enable Save as default Select to enable display of the Save as default control.
Standby Mode timeout Specify the default timeout used when entering Standby
Mode in the Continuous Monitoring profile.
Power line frequency Specify power line frequency for AC power supplied to
the device.
Available profiles Specify the profiles to be available for selection.
Allow profile change Select to enable selection of different profiles. When
disabled, the current profile selection in the Settings tab
is locked in. No other profile selection buttons are
available.
Default profile Specify the default profile to be used at startup in the
following Continuous Monitoring profile workflows: 1)
patient trend data has not been saved, and 2) patient
trend data has been saved and the user selects a new
patient.
4. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Set and start the demo mode
1. Access the Advanced Settings.
a. Touch the Settings tab.
138 Advanced settings Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™