These controls are on the Alarms tab (on the General
tab).
Audio pause time Specify the amount of pause time that is added to the
60-second pause time. When a clinician pauses an audio
alarm tone, the tone is paused for the combined amount
of time.
Nurse call threshold Select the minimum priority alarm that activates a nurse
call relay. If you select High, only high-level alarms
activate a nurse call relay.
SpO2 alarm condition delay Specify the minimum amount of time that an SpO2 alarm
condition must be active before audio and visual signals
occur.
SatSeconds is available with Nellcor SpO2 sensors. If
you select 0 seconds or 10 seconds, SatSeconds is
disabled, and it is removed from the SpO2 tab in the
Alarms tab.
SpHb alarm condition delay Specify the minimum amount of time that an SpHb alarm
condition must be active before audio and visual signals
occur.
etCO2 alarm condition delay Specify the minimum amount of time that an etCO2
alarm condition must be active before audio and visual
signals occur.
Respiration alarm condition delay Specify the minimum amount of time that an RR alarm
condition must be active before audio and visual signals
occur.
No breath detected alarm delay If configured with a CO2 module, specify the amount of
time for the device to wait after receiving a "valid
breath" message before activating the "No breath
detected" physiological alarm.
If configured for RRa, specify the time that the Masimo
module will wait before sending a respiration pause
event, which will result in a "No breath detected"
physiological alarm.
4. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced display settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Display tab.
3. Specify settings.
Setting Action/Description
Display lock Specify the required period of clinician inactivity before
the touchscreen locks.
Directions for use Advanced settings 137