Product Manual
• To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify date and time settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. On the General tab, touch the Date / Time tab.
3. Specify settings.
Setting Action/Description
Date format Select a date format for display.
Time zone Select your time zone offset from Coordinated Universal
Time (UTC).
Automatically adjust clock for daylight saving time,
reported by host
Select this to adjust the displayed time by +/- one hour
when the connected host reports daylight savings time.
Allow users to change date and time Select this to allow clinicians to set the date and time
from the Settings tab.
Display date and time Select this to display the date and time on the Home tab
in the Device Status area.
4. Do one of the following:
• To continue in the Advanced Settings, touch another tab.
• To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced alarm settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Alarms tab.
3. Specify settings.
Option Description
Allow user to disable alarms Select to allow clinicians to turn off or turn on all alarm
limits for each vital sign. The control is on each
parameter-specific tab on the Alarms tab.
Allow user to turn off general audio Select to allow clinicians to turn off all audio notification
for alarms. This control is on the Alarms tab (on the
General tab).
Minimum alarm volume Select the minimum alarm volume available. If you select
High, then Medium and Low are not available to the
clinician.
136 Advanced settings Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™