Product Manual

SpHb numeric view
The numeric view indicates the total hemoglobin level in either grams per deciliter (g/dL)
or millimoles per liter (mmol/L). You can select the unit of measure in Advanced settings.
The size of the SpHb frame and displayed measurements varies based on your
configuration.
Averaging
The averaging button enables you to select the moving window of time used by the
parameter to calculate the SpHb value and update the display: short (approximately 1
minute), medium (approximately 3 minutes), or long (approximately 6 minutes).
SpHb graphical trend view
The graphical trend view presents a trend of the real-time measurements over a user-
selected period. In Advanced settings, you can select the period displayed. The size of
the SpHb frame and displayed trend varies based on your configuration.
The graph shows total hemoglobin level on the y-axis and time on the x-axis (oldest
measurements on the left to newest measurements on the right). The entire graph
updates every 10 seconds.
To the right of the graph, the frame displays the current measurement in numeric
format.
Set up SpHb
Follow these steps to configure SpHb parameter settings.
1.
Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the SpHb vertical tab.
4. Select the desired Trend period.
5. Touch the Home tab.
The new settings take effect immediately.
Configure SpHb alarms
Follow these steps to set alarm limits for SpHb measurements.
1.
Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the Alarms tab.
Directions for use Patient monitoring 121