Product Manual
SpO2 waveform view
The waveform view shows the real-time SpO2 plethysmograph waveform. You can
select the waveform sweep speed in Advanced settings.
Set up SpO2
Follow these steps to configure SpO2 parameter settings.
1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the SpO2 vertical tab.
4. Select the desired SpO2 measurement site using the list box.
5. Select the desired O2 flow rate and O2 concentration using the keypad.
6. Select the desired Sweep speed.
7. Scroll down and select the desired O2 Method from the list box.
8. Touch the Home tab.
The new settings take effect immediately.
Configure SpO2 alarms
Follow these steps to set alarm limits for SpO2 measurements.
1.
Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the Alarms tab.
3. Touch the SpO2 vertical tab.
4. Ensure that the SpO2 alarm limit control is set to ON.
Note If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Enter the desired upper and lower alarms limits for SpO2 using the up/down arrow
keys or the keypad.
6. If your monitor is configured with a Nellcor SpO2 sensor, touch to
select a SatSeconds setting.
7. Touch the Home tab.
The new alarm settings take effect immediately.
Directions for use Patient monitoring 117