Product Manual

8. Return the thermometer to the accessory dock.
The LED on the dock flashes while the measurement is being transferred.
After the transfer is complete, the temperature and the temperature scale appear on
the Home tab according to the monitor settings.
Note Only the latest measurement is transferred to the monitor.
Note Measurements that have already been transferred to the
monitor cannot be transferred again.
For more information about thermometer functionality, refer to the thermometer
manufacturer's directions for use.
Change the temperature scale on the ear thermometer
To switch from Celsius to Fahrenheit, refer to the thermometer manufacturer's
directions for use.
Charge the ear thermometer battery
To charge the battery pack:
Place the thermometer in the accessory dock.
Make sure that the monitor is connected to AC power.
Make sure that the monitor is powered on.
The LED on the dock indicates the charging status of the battery pack:
Orange: The battery pack is charging.
Green: The battery pack is charged.
Not illuminated: The battery pack is not charging.
Note The battery pack continues to charge while the monitor is in Display power
saving mode.
Note It is strongly recommended that you use only the Welch Allyn rechargeable
battery pack in the thermometer because the dock cannot charge other
batteries.
SpO2
SpO2 and pulse rate monitoring continuously measures saturation level of oxygen in
hemoglobin as well as the pulse rate in a patient through a pulse oximeter.
SpO2 frame
The SpO2 frame displays data and controls used in pulse oximetry measurements.
The frame provides a numeric view and a waveform view of SpO2 data. You can toggle
between views by touching the left side of the frame.
SpO2 numeric view
The numeric view indicates the SpO2 saturation percentage and the pulse amplitude.
Features of this view differ based on the type of sensor enabled and the profile selected.
114
Patient monitoring Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™