Product Manual

Note Intervals are not available in all profiles. See the Profiles
section for more information.
Program intervals
You can configure the monitor to take automatic NIBP measurements at variable
intervals. The monitor comes with preset interval programs that can be edited to meet
your needs. You can rename selected programs using the keyboard feature. The
columns below the selected program name indicate the length of each interval in the
cycle (Interval) and the number of intervals you set (Frequency).
Start program intervals
Follow these steps to configure the monitor to take automatic NIBP measurements at
variable intervals.
1.
Properly size the blood pressure cuff and position it around the patient’s bare upper
arm.
2. On the Home tab, touch .
3. Select Program.
4. Touch the desired program.
5. Touch Start intervals.
Create a new program interval or edit an existing program
Follow these steps to create or edit a program interval.
1.
On the Home tab, touch the interval button ( or ).
2. Select Program.
3. Touch the desired program.
4. Touch the keyboard icon and enter the desired program name.
5. Enter the desired interval and frequency settings.
6. Touch Start Intervals.
Directions for use Patient monitoring 101