Product Manual

Take a manual NIBP measurement
WARNING Patient injury risk. Never install Luer Lock connectors on Welch
Allyn blood pressure cuff tubing. Using these connectors on blood pressure
cuff tubing creates the risk of mistakenly connecting this tubing to a
patient's intravenous line and introducing air into the patient's circulatory
system.
CAUTION Inaccurate measurement risk. Any external compression of the
blood pressure hose or cuff may cause system errors or inaccurate
measurements.
1. Properly size the blood pressure cuff and position it around the patient’s bare upper
arm.
2. Touch Start to take a measurement.
Cancel a measurement that is in progress
Follow these steps to cancel any NIBP measurement that is in progress.
On the Home tab, touch .
The monitor rapidly deflates the cuff, and the screen displays the NIBP cancellation
message.
Interval NIBP measurement
The monitor can take NIBP measurements automatically based on intervals you choose.
The Intervals tab provides all interval features.
From this tab, you can do the following:
Configure intervals
Turn off intervals
Configure the monitor to print automatic measurements as they are completed
When the measurement is complete, the NIBP frame displays the measurement until
the next measurement is complete.
Note During intervals, each automatic and manual save of patient measurements
clears all measurements from Manual parameters frame.
The button changes to a timer ( ), which counts down to the next
automatic measurement.
Automatic measurements continue until you turn off intervals.
WARNING Patient harm risk. Do not use intervals on neonates out of
earshot. Verify that audio can be heard from where you intend to be.
Directions for use Patient monitoring 99