Product Manual

Note In situations where you cannot position the cuff level with
the heart, you should adjust the measurements as follows
for greater accuracy. For each inch (2.54 cm) that the cuff is
above the level of the heart, add 1.8 mmHg to the displayed
reading. For each inch (2.54 cm) that the cuff is below the
level of the heart, subtract 1.8 mmHg from the displayed
reading. It is important to document the adjustment on the
patient record.
Set up NIBP
Follow these steps to configure NIBP parameter settings.
1. Touch the Settings tab.
2. Touch the Setup tab.
3. Touch the NIBP vertical tab.
4. Adjust the following settings as desired:
Cuff site. Select the site from the list box.
Cuff size. Select the size from the list box.
Patient position.Select the position from the list box.
Cuff inflation target. If the monitor is set up for StepBP, enter the Cuff inflation
target. You can set the default in Advanced settings.
5. Touch the Home tab.
The new settings take effect immediately.
Configure NIBP alarms
Follow these steps to set alarm limits for systolic and diastolic measurements, and MAP
calculation.
1.
Verify that you are using the Intervals Monitoring or Continuous Monitoring profile.
2. Touch the Alarms tab.
3. Touch the NIBP vertical tab.
4. Ensure that the NIBP and MAP alarm limit controls are set to ON.
Note If any parameter's alarm limit control is set to OFF, you
cannot adjust alarm limits on the Alarm tab, and no visual or
audio signals will occur for that specific parameter.
5. Adjust the following settings as desired:
Systolic. Enter upper and lower Systolic alarm limits using the up/down arrow
keys or the keypad.
Diastolic. Enter upper and lower Diastolic alarm limits using the up/down arrow
keys or the keypad.
MAP. Enter upper and lower MAP alarm limits using the up/down arrow keys or
the keypad.
6. Touch the Home tab.
Directions for use Patient monitoring 97