Product Manual

Buttons
The buttons on the right side of the frame enable you to do different tasks depending on
the profile you are using. The availability of functions depends on which profile is
selected. See the Profiles section for more information.
Button name Button image Description
Start/Stop
The appearance and function of this button dynamically changes.
Touch to start a manual measurement or a cycle of
automatic measurements.
Touch to stop a measurement that is in progress.
Interval
This button shows the status of automatic measurements.
Touch the button to display the Intervals tab, where you can configure automatic
measurements.
Automatic measurements are off.
Automatic measurements are on.
Alarm Limit control This button displays alarm status. It also displays alarm limits in some device
configurations.
Touch the button to display the Alarms tab.
Select a cuff
WARNING Use only blood pressure cuffs and hoses listed as approved
accessories to ensure safe and accurate NIBP measurements.
WARNING Never use an adult or pediatric monitor setting or cuff for an
NIBP measurement on a neonatal patient. Adult and pediatric inflation
limits can be excessive for neonatal patients, even if a neonatal cuff is
used. Neonates are defined in the AAMI SP10:2002 standard as children 28
days or less of age if born at term (37 weeks gestation or more); otherwise,
up to 44 gestational weeks.
94 Patient monitoring Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™