2 ENGLISH Cleaning, Disinfection, and Sterilization Instructions
Rigid Sigmoidoscopes, Anoscopes, and Light Carrier
*Suction Tubes not illustrated
Cleaning
1. Prepare an enzymatic detergent formulated for endoscopic instruments according to
the manufacturer’s instructions.These instructions were validated using ENZOL
®
/
CIDEZYME
®
.
2. Submerge the item in the cleaning solution.
3. Using a soft bristle brush, scrub the item submerged in the cleaning solution for a
minimum of 5 minutes then rinse with sterile water for 5 minutes.
High Level Disinfection
1. Follow the Cleaning instructions above.
2. Using CIDEX
®
OPA (0.55% OPA, 14-day maximum reuse), soak the item for a
minimum of 12 minutes.
3. Perform 3 sequential rinses of the item for a minimum of 5 minutes each with sterile
water.
Anoscope
Part Numbers:
Sigmoidoscope
Part Numbers:
Light Carrier
Part Numbers:
37019 38614 32010 32820 36019
37023 38619 32020 32830 38700
37027 38622 32410 33220 Suction Tube*
Part Numbers:
38108 38900 32420 33620
38114 39614 32810 33830 30130
38119 39619 30140
38122 39622
Note
For all steps, remove obturator for separate cleaning and open viewing window.
Caution Welch Allyn recommends the use of the Single Use Insufflation Bulb
Filter ( 30210) for all applicable endoscopic procedures. The filter does not
eliminate the need to properly process the insufflation bulb with each use, but
restricts internal contamination and aspiration of fluids.