Product Manual

Rigid Reusable & Single use Sigmoidoscopes, Anoscopes, and Accessories ENGLISH 9
Rechargeable Handle
Cleaning
1. Using CaviWipes, wipe the surface to remove all gross debris.
2. Using a soft bristle brush, scrub the handle for a minimum of 5 minutes then repeat
wipe with a second CaviWipe.
3. Allow the handle surface to dry for a minimum of 10 minutes.
Intermediate Level Disinfection
1. Follow the Cleaning instructions above.
2. Using CaviWipes, wipe the handle and keep wet with CaviWipe solution for a
minimum of 5 minutes.
3. Allow the handle surface to dry for a minimum of 10 minutes.
Rheostat
(Brightness
Control)
Part Number 71000-A
Caution Do not saturate the rheostat area. Excess solution entering the
assembly may damageinternal components in the handle.